SYSTEM DESCRIPTION
Open, random access, counter-top instrument for chemical and immunoturbidimetric clinical analysis.
METHODOLOGIES
Kinetic, fixed time, end point, bichromatic, differential, unlimited different methods can be programmed.
OPERATIVE MODE
– Profile
– Batch
– Selective
– STAT
THROUGHPUT
Up to 138 tests/hour.
SAMPLE LOADING
– Continuous loading on 4 separate racks, 8 positions each.
– STAT samples can be placed at any time in any of the 4 racks and be run immediately
SAMPLE CONTAINERS
– Primary tubes diameter from 8 to 16 mm; height up to 100 mm.
– Cups from 1 ml to 3 ml
SAMPLING
– Sample volume: 2.0 to 99 microliters (1 ? increments)
– Reagent volume: 3.0 to 548 microliters (1 ? increments)
– Capacitive level sensor
– Reagent preheating to 37 °C
– Automatic sample probe washing (internal and external)
– Probe wash cycle is operator programmable
REAGENTS
24 positions available for reagent containers.
STANDARDS AND CONTROLS
Up to 12 positions.
MINIMUM REACTION VOLUME
200 microliters.
READING CUVETTES
120 semi – disposable, optic grade cuvettes divided among 6 racks (20 each).
REACTION TEMPERATURE
Air bath at 37°C + 0.2.
OPTIC MODULE
– Double beam
– Up to 9 interferential filters + 1 position available for an optional filter
– Narrow band filters 340, 380, 405, 492, 510, 546, 578, 620, 690 (optional)
– FWHM: 10nm + 2nm
– 6V/10W halogen bulb
– Optic path length: 6 mm
– Resolution: 0,0005 Abs
– Linearity: up to 4,100 Abs
OPTIONAL ACCESSORIES
Bar code reader (reagents and samples)